Smiths-Medical Medfusion 4000 Wireless Syringe Infusion Pump 1.1 Network Configuration hard-coded credentials
CVSS Meta Temp Score | Current Exploit Price (≈) | CTI Interest Score |
---|---|---|
6.5 | $0-$5k | 0.00 |
A vulnerability has been found in Smiths-Medical Medfusion 4000 Wireless Syringe Infusion Pump 1.1 (Medical Device Software) and classified as critical. This vulnerability affects some unknown processing of the component Network Configuration. The manipulation with an unknown input leads to a hard-coded credentials vulnerability. The CWE definition for the vulnerability is CWE-798. The product contains hard-coded credentials, such as a password or cryptographic key, which it uses for its own inbound authentication, outbound communication to external components, or encryption of internal data. As an impact it is known to affect confidentiality, integrity, and availability. CVE summarizes:
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
The weakness was published 02/15/2018 by Scott Gayou (Website). The advisory is available at securityfocus.com. This vulnerability was named CVE-2017-12725 since 08/09/2017. The attack can be initiated remotely. No form of authentication is required for a successful exploitation. The technical details are unknown and an exploit is not available. This vulnerability is assigned to T1110.001 by the MITRE ATT&CK project.
There is no information about possible countermeasures known. It may be suggested to replace the affected object with an alternative product.
Similar entries are available at 113344, 113345, 113346 and 113347.
Product
Type
Vendor
Name
Version
CPE 2.3
CPE 2.2
CVSSv4
VulDB CVSS-B Score: 🔍VulDB CVSS-BT Score: 🔍
VulDB Vector: 🔍
VulDB Reliability: 🔍
CVSSv3
VulDB Meta Base Score: 6.5VulDB Meta Temp Score: 6.5
VulDB Base Score: 7.3
VulDB Temp Score: 7.3
VulDB Vector: 🔍
VulDB Reliability: 🔍
NVD Base Score: 5.6
NVD Vector: 🔍
CVSSv2
AV | AC | Au | C | I | A |
---|---|---|---|---|---|
💳 | 💳 | 💳 | 💳 | 💳 | 💳 |
💳 | 💳 | 💳 | 💳 | 💳 | 💳 |
💳 | 💳 | 💳 | 💳 | 💳 | 💳 |
Vector | Complexity | Authentication | Confidentiality | Integrity | Availability |
---|---|---|---|---|---|
unlock | unlock | unlock | unlock | unlock | unlock |
unlock | unlock | unlock | unlock | unlock | unlock |
unlock | unlock | unlock | unlock | unlock | unlock |
VulDB Base Score: 🔍
VulDB Temp Score: 🔍
VulDB Reliability: 🔍
NVD Base Score: 🔍
Exploiting
Class: Hard-coded credentialsCWE: CWE-798 / CWE-259 / CWE-255
ATT&CK: T1110.001
Local: No
Remote: Yes
Availability: 🔍
Status: Not defined
EPSS Score: 🔍
EPSS Percentile: 🔍
Price Prediction: 🔍
Current Price Estimation: 🔍
0-Day | unlock | unlock | unlock | unlock |
---|---|---|---|---|
Today | unlock | unlock | unlock | unlock |
Threat Intelligence
Interest: 🔍Active Actors: 🔍
Active APT Groups: 🔍
Countermeasures
Recommended: no mitigation knownStatus: 🔍
0-Day Time: 🔍
Timeline
08/09/2017 🔍09/07/2017 🔍
02/15/2018 🔍
02/15/2018 🔍
02/16/2018 🔍
01/05/2020 🔍
Sources
Advisory: securityfocus.com⛔Researcher: Scott Gayou
Status: Not defined
CVE: CVE-2017-12725 (🔍)
SecurityFocus: 100665 - Medfusion 4000 Wireless Syringe Infusion Pump ICSMA-17-250-02 Multiple Security Vulnerabilities
See also: 🔍
Entry
Created: 02/16/2018 08:37Updated: 01/05/2020 11:58
Changes: 02/16/2018 08:37 (60), 01/05/2020 11:58 (3)