Baxter SIGMA Spectrum Infusion System 6.05 Wireless Battery Module improper authentication
CVSS Meta Temp Score | Current Exploit Price (≈) | CTI Interest Score |
---|---|---|
8.4 | $0-$5k | 0.00 |
A vulnerability was found in Baxter SIGMA Spectrum Infusion System 6.05 (Medical Device Software). It has been declared as critical. This vulnerability affects an unknown code block of the component Wireless Battery Module. The manipulation with an unknown input leads to a improper authentication vulnerability. The CWE definition for the vulnerability is CWE-287. When an actor claims to have a given identity, the product does not prove or insufficiently proves that the claim is correct. As an impact it is known to affect confidentiality, integrity, and availability. CVE summarizes:
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
The weakness was shared 03/26/2019. This vulnerability was named CVE-2014-5432 since 08/22/2014. The exploitation appears to be easy. The attack can be initiated remotely. No form of authentication is required for a successful exploitation. There are neither technical details nor an exploit publicly available.
Upgrading to version 8 eliminates this vulnerability.Proper firewalling of tcp/22 (ssh) is able to address this issue. The best possible mitigation is suggested to be upgrading to the latest version.
The entries 132426, 132425 and 132423 are related to this item.
Product
Type
Vendor
Name
Version
CPE 2.3
CPE 2.2
CVSSv4
VulDB CVSS-B Score: 🔍VulDB CVSS-BT Score: 🔍
VulDB Vector: 🔍
VulDB Reliability: 🔍
CVSSv3
VulDB Meta Base Score: 8.5VulDB Meta Temp Score: 8.4
VulDB Base Score: 7.3
VulDB Temp Score: 7.0
VulDB Vector: 🔍
VulDB Reliability: 🔍
NVD Base Score: 9.8
NVD Vector: 🔍
CVSSv2
AV | AC | Au | C | I | A |
---|---|---|---|---|---|
💳 | 💳 | 💳 | 💳 | 💳 | 💳 |
💳 | 💳 | 💳 | 💳 | 💳 | 💳 |
💳 | 💳 | 💳 | 💳 | 💳 | 💳 |
Vector | Complexity | Authentication | Confidentiality | Integrity | Availability |
---|---|---|---|---|---|
unlock | unlock | unlock | unlock | unlock | unlock |
unlock | unlock | unlock | unlock | unlock | unlock |
unlock | unlock | unlock | unlock | unlock | unlock |
VulDB Base Score: 🔍
VulDB Temp Score: 🔍
VulDB Reliability: 🔍
NVD Base Score: 🔍
Exploiting
Class: Improper authenticationCWE: CWE-287
CAPEC: 🔍
ATT&CK: 🔍
Local: No
Remote: Yes
Availability: 🔍
Status: Not defined
EPSS Score: 🔍
EPSS Percentile: 🔍
Price Prediction: 🔍
Current Price Estimation: 🔍
0-Day | unlock | unlock | unlock | unlock |
---|---|---|---|---|
Today | unlock | unlock | unlock | unlock |
Threat Intelligence
Interest: 🔍Active Actors: 🔍
Active APT Groups: 🔍
Countermeasures
Recommended: UpgradeStatus: 🔍
0-Day Time: 🔍
Upgrade: SIGMA Spectrum Infusion System 8
Firewalling: 🔍
Timeline
08/22/2014 🔍03/26/2019 🔍
03/27/2019 🔍
08/08/2023 🔍
Sources
Advisory: ics-cert.us-cert.govStatus: Not defined
CVE: CVE-2014-5432 (🔍)
See also: 🔍
Entry
Created: 03/27/2019 06:35Updated: 08/08/2023 12:11
Changes: 03/27/2019 06:35 (39), 05/21/2020 15:39 (18), 08/08/2023 12:11 (4)
Complete: 🔍
No comments yet. Languages: en.
Please log in to comment.