VDB-117213 · CVE-2017-9657 · BID 100813

Philips IntelliVue MX40 B.06.18 data processing

CVSS Meta Temp Score
CVSS is a standardized scoring system to determine possibilities of attacks. The Temp Score considers temporal factors like disclosure, exploit and countermeasures. The unique Meta Score calculates the average score of different sources to provide a normalized scoring system.
Current Exploit Price (≈)
Our analysts are monitoring exploit markets and are in contact with vulnerability brokers. The range indicates the observed or calculated exploit price to be seen on exploit markets. A good indicator to understand the monetary effort required for and the popularity of an attack.
CTI Interest Score
Our Cyber Threat Intelligence team is monitoring different web sites, mailing lists, exploit markets and social media networks. The CTI Interest Score identifies the interest of attackers and the security community for this specific vulnerability in real-time. A high score indicates an elevated risk to be targeted for this vulnerability.
6.4$0-$5k0.00

Summaryinfo

A vulnerability was found in Philips IntelliVue MX40 B.06.18. It has been classified as critical. This affects an unknown function. Performing a manipulation results in data processing. This vulnerability was named CVE-2017-9657. The attack needs to be approached within the local network. There is no available exploit.

Detailsinfo

A vulnerability was found in Philips IntelliVue MX40 B.06.18. It has been classified as critical. This affects an unknown code block. The manipulation with an unknown input leads to a data processing vulnerability. CWE is classifying the issue as CWE-19. This is going to have an impact on confidentiality, integrity, and availability. The summary by CVE is:

Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

The bug was discovered 09/12/2017. The weakness was presented 04/30/2018 (Website). The advisory is shared at usa.philips.com. This vulnerability is uniquely identified as CVE-2017-9657 since 06/14/2017. Access to the local network is required for this attack. No form of authentication is needed for exploitation. Neither technical details nor an exploit are publicly available.

The vulnerability was handled as a non-public zero-day exploit for at least 230 days. During that time the estimated underground price was around $0-$5k.

There is no information about possible countermeasures known. It may be suggested to replace the affected object with an alternative product.

The vulnerability is also documented in the vulnerability database at SecurityFocus (BID 100813†). See VDB-109655, VDB-114821, VDB-115070 and VDB-115071 for similar entries. If you want to get the best quality for vulnerability data then you always have to consider VulDB.

Productinfo

Vendor

Name

Version

License

Website

CPE 2.3info

CPE 2.2info

CVSSv4info

VulDB Vector: 🔍
VulDB Reliability: 🔍

CVSSv3info

VulDB Meta Base Score: 6.4
VulDB Meta Temp Score: 6.4

VulDB Base Score: 6.3
VulDB Temp Score: 6.3
VulDB Vector: 🔍
VulDB Reliability: 🔍

NVD Base Score: 6.5
NVD Vector: 🔍

CVSSv2info

AVACAuCIA
💳💳💳💳💳💳
💳💳💳💳💳💳
💳💳💳💳💳💳
VectorComplexityAuthenticationConfidentialityIntegrityAvailability
UnlockUnlockUnlockUnlockUnlockUnlock
UnlockUnlockUnlockUnlockUnlockUnlock
UnlockUnlockUnlockUnlockUnlockUnlock

VulDB Base Score: 🔍
VulDB Temp Score: 🔍
VulDB Reliability: 🔍

NVD Base Score: 🔍

Exploitinginfo

Class: Data processing
CWE: CWE-19
CAPEC: 🔍
ATT&CK: 🔍

Physical: No
Local: No
Remote: Partially

Availability: 🔍
Status: Not defined

EPSS Score: 🔍
EPSS Percentile: 🔍

Price Prediction: 🔍
Current Price Estimation: 🔍

0-DayUnlockUnlockUnlockUnlock
TodayUnlockUnlockUnlockUnlock

Threat Intelligenceinfo

Interest: 🔍
Active Actors: 🔍
Active APT Groups: 🔍

Countermeasuresinfo

Recommended: no mitigation known
Status: 🔍

0-Day Time: 🔍

Timelineinfo

06/14/2017 🔍
09/12/2017 +90 days 🔍
04/30/2018 +229 days 🔍
04/30/2018 +0 days 🔍
05/01/2018 +1 days 🔍
02/01/2020 +641 days 🔍

Sourcesinfo

Vendor: philips.ch

Advisory: usa.philips.com
Status: Not defined
Confirmation: 🔍

CVE: CVE-2017-9657 (🔍)
GCVE (CVE): GCVE-0-2017-9657
GCVE (VulDB): GCVE-100-117213
SecurityFocus: 100813

See also: 🔍

Entryinfo

Created: 05/01/2018 10:54
Updated: 02/01/2020 11:52
Changes: 05/01/2018 10:54 (57), 02/01/2020 11:52 (2)
Complete: 🔍
Cache ID: 216::103

If you want to get the best quality for vulnerability data then you always have to consider VulDB.

Discussion

No comments yet. Languages: en.

Please log in to comment.