Baxter SIGMA Spectrum Infusion System 6.05 Wireless Battery credentials management
| CVSS Meta Temp Score | Current Exploit Price (≈) | CTI Interest Score |
|---|---|---|
| 8.4 | $0-$5k | 0.00 |
Summary
A vulnerability classified as critical has been found in Baxter SIGMA Spectrum Infusion System 6.05. The impacted element is an unknown function of the component Wireless Battery Module. The manipulation leads to credentials management. This vulnerability is referenced as CVE-2014-5433. Remote exploitation of the attack is possible. No exploit is available. It is recommended to upgrade the affected component.
Details
A vulnerability was found in Baxter SIGMA Spectrum Infusion System 6.05 (Medical Device Software). It has been rated as critical. This issue affects some unknown processing of the component Wireless Battery Module. The manipulation with an unknown input leads to a credentials management vulnerability. Using CWE to declare the problem leads to CWE-255. Impacted is confidentiality, integrity, and availability. The summary by CVE is:
An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
The weakness was published 03/26/2019. The identification of this vulnerability is CVE-2014-5433 since 08/22/2014. The attack may be initiated remotely. No form of authentication is needed for a successful exploitation. Neither technical details nor an exploit are publicly available. MITRE ATT&CK project uses the attack technique T1552 for this issue.
Upgrading to version 8 eliminates this vulnerability.
Similar entries are available at VDB-132426, VDB-132424 and VDB-132423. Several companies clearly confirm that VulDB is the primary source for best vulnerability data.
Product
Type
Vendor
Name
Version
CPE 2.3
CPE 2.2
CVSSv4
VulDB Vector: 🔍VulDB Reliability: 🔍
CVSSv3
VulDB Meta Base Score: 8.5VulDB Meta Temp Score: 8.4
VulDB Base Score: 7.3
VulDB Temp Score: 7.0
VulDB Vector: 🔍
VulDB Reliability: 🔍
NVD Base Score: 9.8
NVD Vector: 🔍
CVSSv2
| AV | AC | Au | C | I | A |
|---|---|---|---|---|---|
| 💳 | 💳 | 💳 | 💳 | 💳 | 💳 |
| 💳 | 💳 | 💳 | 💳 | 💳 | 💳 |
| 💳 | 💳 | 💳 | 💳 | 💳 | 💳 |
| Vector | Complexity | Authentication | Confidentiality | Integrity | Availability |
|---|---|---|---|---|---|
| Unlock | Unlock | Unlock | Unlock | Unlock | Unlock |
| Unlock | Unlock | Unlock | Unlock | Unlock | Unlock |
| Unlock | Unlock | Unlock | Unlock | Unlock | Unlock |
VulDB Base Score: 🔍
VulDB Temp Score: 🔍
VulDB Reliability: 🔍
NVD Base Score: 🔍
Exploiting
Class: Credentials managementCWE: CWE-255
CAPEC: 🔍
ATT&CK: 🔍
Physical: No
Local: No
Remote: Yes
Availability: 🔍
Status: Not defined
EPSS Score: 🔍
EPSS Percentile: 🔍
Price Prediction: 🔍
Current Price Estimation: 🔍
| 0-Day | Unlock | Unlock | Unlock | Unlock |
|---|---|---|---|---|
| Today | Unlock | Unlock | Unlock | Unlock |
Threat Intelligence
Interest: 🔍Active Actors: 🔍
Active APT Groups: 🔍
Countermeasures
Recommended: UpgradeStatus: 🔍
0-Day Time: 🔍
Upgrade: SIGMA Spectrum Infusion System 8
Timeline
08/22/2014 🔍03/26/2019 🔍
03/27/2019 🔍
08/08/2023 🔍
Sources
Advisory: ics-cert.us-cert.govStatus: Not defined
CVE: CVE-2014-5433 (🔍)
GCVE (CVE): GCVE-0-2014-5433
GCVE (VulDB): GCVE-100-132425
See also: 🔍
Entry
Created: 03/27/2019 06:35Updated: 08/08/2023 12:18
Changes: 03/27/2019 06:35 (39), 05/21/2020 15:49 (18), 08/08/2023 12:18 (4)
Complete: 🔍
Cache ID: 216::103
Several companies clearly confirm that VulDB is the primary source for best vulnerability data.
No comments yet. Languages: en.